Key Features Of The UID-HSR USB Interface Device
Connects Analog Sensors To A PC
Accepts one 4-20 mA signal or one 0-10 V signal, single ended or differential.PT100 Temperature Input
A second channel for a Class B PT100 sensor, 3-wire or 4-wire, powered by the interface.Up To 1000 Samples Per Second
A user-selectable sample rate for both slow trends and fast events, with ±0.15% of span accuracy.High Resolution
21 bits at sample rates up to 5 Hz and 16 bits from 5 Hz to 1000 Hz.USB Powered With Auto-Detection
The interface takes its power from the USB port and is recognized automatically by the software.IP65 Enclosure With Screw Terminals
An ABS housing with three PG7 cable glands and terminals for 26 to 14 AWG wire.
Applications For USB Sensor Interfaces
The UID-HSR is useful wherever a 4-20 mA, 0-10 V or PT100 signal needs to be viewed or recorded on a computer:
- Laboratory testing – logging pressure, flow, level or force sensors alongside temperature.
- Process monitoring – temporary or permanent recording of plant signals on a PC.
- Leak testing – automated tests with reports and certificates.
- Commissioning and fault finding – a portable way to record a transmitter signal on a laptop.
- Sensor checking and calibration – comparing a sensor output against a reference.
- Research and development – adding PC data logging to existing analog sensors.
How The UID-HSR Turns Any Sensor Into A USB Data Logger
Most industrial sensors output a 4-20 mA or 0-10 V signal intended for a PLC or panel meter. To record that signal on a computer you would normally need a data acquisition card, a power supply and software to scale the readings. The UID-HSR replaces these with a single USB device.
Wire the sensor to the screw terminals, connect the USB cable and the software detects the interface automatically. Readings can then be scaled to any engineering unit, so a 4-20 mA signal can be displayed as bar, psi, liters per minute or any other unit you define. The interface itself is USB powered and also powers the PT100; the connected analog sensor needs its own supply.
PC Software Features
- Graphical and tabular data display with customizable scales
- User-defined engineering units
- Individual alarm levels for each sensor
- Up to 16 sensors managed at the same time
- Virtual differential measurement between sensors
- Leak testing and automatic test scheduling
- Custom certificates and customizable report templates
- Data export to Excel and PDF
Inputs And Connections
| Channel | Signal | Sample Rate |
|---|---|---|
| Analog input | 4-20 mA (single ended) or 0-10 VDC (single ended or differential) | Up to 1000 samples per second |
| Temperature input | PT100 Class B, 3-wire or 4-wire | Up to 5 samples per second |
UID-HSR USB Interface Device Technical Specifications
| Specification | Detail |
|---|---|
| Product Type | USB interface device for sensor data logging |
| Analog Input | 1 channel: 4-20 mA single ended, or 0-10 VDC single ended or differential |
| Temperature Input | 1 channel: PT100 Class B (IEC 60751), 3-wire or 4-wire |
| Accuracy | ±0.15% of span BFSL |
| Sample Rate | User selectable up to 1000 samples per second; temperature 5 samples per second maximum |
| Resolution | 21 bits at ≤ 5 Hz; 16 bits from 5 Hz to 1000 Hz |
| Interface | USB 2.0, compatible with USB 3.x |
| Power Supply | 5 V from the USB port; external supply required for connected sensors |
| USB Connection | USB mini B socket with 2 m lead to USB type A |
| Sensor Connections | 3 x PG7 cable glands; screw terminals for 26 to 14 AWG wire |
| Enclosure | ABS plastic, 158 x 90 x 45 mm, six 7 mm mounting holes |
| Ingress Protection | IP65 |
| Weight | 300 g |
| Operating Temperature | -20 to +80 °C (-4 to +176 °F) |
| Storage Temperature | +5 to +40 °C (+41 to +104 °F) |
| Operating Systems | Windows 8, 10 and 11 |
| Sensors Per PC | Up to 16 managed simultaneously in the software |
UID-HSR USB Interface Device FAQs
Which sensors can I connect to the UID-HSR?
Any sensor or transmitter with a 4-20 mA or 0-10 V output, such as pressure, level, flow, force or position sensors, plus a PT100 temperature sensor.
Does the UID-HSR power the connected sensor?
The interface is powered from the USB port and powers the PT100 temperature sensor. A 4-20 mA or 0-10 V sensor needs its own external power supply.
How many sensors can I log at once?
Each UID-HSR has one analog input and one PT100 input. The software can manage up to 16 sensors at the same time using multiple devices.
Can readings be shown in my own units?
Yes. The software scales the input signal to any user-defined engineering unit.