The Short Answer
A magnetic flowmeter creates a magnetic field across the pipe. When a conductive liquid flows through the field, it generates a small voltage that is directly proportional to its velocity. Electrodes pick up this voltage, and the meter converts it into a flow rate. There are no moving parts and nothing obstructs the flow.
Faraday’s Law in Action
Faraday’s law of induction states that a conductor moving through a magnetic field generates a voltage. In a mag meter, the conductor is the liquid itself. Coils on the outside of the flow tube create the magnetic field, and two electrodes in the pipe wall measure the induced voltage.
Induced voltage = Magnetic field strength × Flow velocity × Pipe diameter
Because the field strength and diameter are fixed, the voltage depends only on velocity. Multiply velocity by the pipe area and you have the volumetric flow rate.
What Mag Meters Can Measure
- Conductive liquids, typically above 5 µS/cm: water, wastewater, slurries, acids and alkalis.
- Dirty liquids and slurries, since there is nothing to clog.
- Not hydrocarbons, oils, gases, steam or deionized water, which are not conductive enough.
Advantages and Limitations
Advantages
- No moving parts and no pressure drop.
- Unaffected by density, viscosity and temperature.
- Accurate across a wide flow range.
- Bidirectional measurement is possible.
Limitations
- Liquid must be conductive.
- Pipe must run full.
- Liner and electrodes must suit the media.
- Needs power for the coils.
Installation Tips
- Provide straight pipe upstream and downstream, typically 5 diameters before and 3 after.
- Install in a vertical pipe with upward flow, or at a low point in horizontal pipe, so it always runs full.
- Ground the meter and liquid correctly, using grounding rings in plastic or lined pipe.
- Avoid locations where air can collect, such as the top of a pipe run.
Related Products
We supply inline magnetic flowmeters and the displays and totalizers to go with them.
Frequently Asked Questions
Can a magnetic flowmeter measure oil?
No. Oil is not electrically conductive, so a mag meter cannot generate a signal. Consider a turbine, gear or ultrasonic meter instead.
Does a mag meter need straight pipe?
Less than most technologies, but a few diameters upstream and downstream improve accuracy.
What is the minimum conductivity?
Most mag meters need at least 5 µS/cm. Some special designs work down to lower values.