How Is Pressure Sensor Accuracy Specified? %FSO, BFSL and Total Error Band Explained

The Short Answer

Pressure sensor accuracy is usually quoted as a percentage of full scale output (%FSO). The headline figure often covers only non-linearity, hysteresis and repeatability at room temperature. The total error band (TEB) adds temperature effects and zero and span offsets, and gives a much better idea of real-world performance.

Why %FSO Matters

Most accuracy figures are a percentage of the full range, not of the reading. A 0-100 psi transmitter with ±0.25% FSO accuracy can be out by ±0.25 psi anywhere in its range. At 10 psi, that is 2.5% of the reading. That is why you should select a range close to your normal operating pressure.

Maximum error = Accuracy (%FSO) ÷ 100 × Full scale range

The Components of Accuracy

BFSL vs Terminal Point Linearity

Non-linearity can be calculated two ways. Best fit straight line (BFSL) draws the line that minimizes the error across the range, and gives a smaller number. Terminal point draws the line between the actual zero and full scale points, and typically gives roughly double the figure for the same sensor. When comparing datasheets, check which method each uses.

Total Error Band

Total error band combines every error source across the full compensated temperature range into one figure. It is the most honest measure of how far a reading can be from the true value in service. A sensor quoted at ±0.1% BFSL may have a total error band of ±1% or more across -20 to +85 °C.

±0.1%

Typical
BFSL accuracy

±0.2%/10 K

Typical
thermal error

±1%

Typical
total error band

How to Get the Best Accuracy

Related Products

Our range runs from cost-effective OEM transducers to high-accuracy transmitters with tight total error bands.

Industrial transmitters with detailed accuracy data. View range

Customizable OEM sensors for volume applications. View range

High-resolution sensors for low ranges. View range

Frequently Asked Questions

It depends where you operate. Near full scale, %FSO and % of reading are similar. At low pressures, % of reading gives a much smaller error.

Most process control needs ±0.25 to ±0.5% FSO. Custody transfer, test benches and calibration work may need ±0.1% or better.

Yes. Long-term stability, often ±0.1 to ±0.2% FSO per year, describes the expected drift, which is why periodic calibration is recommended.

More in This Category

Pressure Sensor vs Transducer vs Transmitter vs Switch: What Is the Difference?

All four measure pressure, but they differ in their output. A pressure sensor is the sensing element itself. A pressure transducer usually gives a low-level voltage output, such as mV/V or 0.5-4.5 V. A pressure transmitter gives a conditioned, standardized signal, typically 4-20 mA, suited to long cable runs. A pressure switch opens or closes a contact when the pressure crosses a set point.

Ceramic vs Piezoresistive vs Thin-Film: Which Pressure Sensing Element?

Ceramic sensors resist corrosion and abrasion and handle overpressure well, making them a strong choice for water, wastewater and chemicals at low to medium pressures. Piezoresistive silicon sensors offer the best accuracy and sensitivity, especially at low ranges. Thin-film sensors, sputtered onto a stainless steel diaphragm, are rugged and stable at high pressures and temperatures, ideal for hydraulics.

Need help choosing the right product?

Our engineers can help you select, configure and integrate the right sensor, instrument or industrial computer for your application.