The Short Answer
Each thermocouple type uses a different pair of metals, which sets its temperature range and best uses. Type K is the general-purpose choice up to about 1,260 °C. Type J suits lower temperatures and reducing atmospheres, Type T is best for low and cryogenic temperatures, and Types R and S cover very high temperatures. Every thermocouple measures relative to its cold junction, so the instrument must apply cold junction compensation.
Common Thermocouple Types
- Type K (chromel-alumel): -200 to 1,260 °C. General purpose, oxidizing atmospheres.
- Type J (iron-constantan): -40 to 760 °C. Reducing atmospheres and older equipment.
- Type T (copper-constantan): -200 to 370 °C. Food, cryogenics and low temperatures.
- Type E (chromel-constantan): -200 to 870 °C. Highest output, good at low temperatures.
- Type N (nicrosil-nisil): -200 to 1,260 °C. More stable than K at high temperatures.
- Types R and S (platinum-rhodium): up to about 1,600 °C. Furnaces and glass.
How a Thermocouple Works
When two different metals are joined and the junction is heated, a small voltage appears across the free ends. This is the Seebeck effect. The voltage is tiny, roughly 41 µV per °C for Type K, and depends on the temperature difference between the hot (measuring) junction and the cold (reference) junction where the wires connect to the instrument.
Cold Junction Compensation
Because the voltage depends on a temperature difference, the instrument must know the temperature at its own terminals. It measures that with a built-in sensor and adds the equivalent voltage back, so the result reflects the true hot junction temperature. This is cold junction compensation (CJC).
Measured temperature = Thermocouple reading + Cold junction temperature
Use the Right Extension Wire
Every connection between dissimilar metals forms a new junction. Run matching thermocouple or extension wire all the way to the instrument, and use thermocouple connectors of the same type. Ordinary copper wire creates extra junctions and errors that change with ambient temperature.
- Match extension wire type to the thermocouple type.
- Observe polarity. In the US (ANSI), the red lead is negative.
- Convert to 4-20 mA with a transmitter close to the sensor for long runs.
Related Products
Our instruments accept all common thermocouple types with automatic cold junction compensation.
Frequently Asked Questions
Which thermocouple type is most common?
Type K, because it covers a wide range at low cost and suits most oxidizing environments.
Why does my thermocouple read room temperature when disconnected?
The instrument is still applying cold junction compensation, so an open circuit may show ambient temperature or trigger a burnout alarm.
Can I use copper wire to extend a thermocouple?
No. Copper creates new junctions that introduce errors. Use matching thermocouple extension wire.