What Is a Pt100 RTD, and What Do Class A and Class B Mean?

The Short Answer

A Pt100 is a platinum resistance temperature detector with a resistance of exactly 100 Ω at 0 °C, rising by about 0.385 Ω per °C. Its tolerance class, defined in IEC 60751, tells you how accurate it is. Class A is ±0.15 °C at 0 °C, and Class B is ±0.3 °C at 0 °C, with the tolerance widening as temperature moves away from zero.

How a Pt100 Works

Platinum is used because its resistance changes in a very stable, repeatable way with temperature. The standard alpha coefficient of 0.00385 means the resistance rises by 0.385% of its 0 °C value for each degree. At 100 °C, a Pt100 reads about 138.5 Ω.

18.5 Ω

Resistance at
-200 °C

100 Ω

Resistance at
0 °C

138.5 Ω

Resistance at
100 °C

247.1 Ω

Resistance at
400 °C

IEC 60751 Tolerance Classes

The tolerance grows with temperature according to a simple formula, where t is the temperature in °C:

Example: at 100 °C, Class A allows ±0.35 °C and Class B allows ±0.8 °C.

Pt100 vs Pt1000

A Pt1000 works the same way but measures 1,000 Ω at 0 °C. Its higher resistance means lead wire resistance causes a smaller error, so Pt1000 sensors suit 2-wire connections and battery-powered devices. Pt100 remains the standard in process industries.

Getting the Most From a Pt100

Related Products

Our transmitters and instruments accept Pt100 and Pt1000 inputs directly.

Pt100 inputs converted to 4-20 mA. View range

Controllers with RTD inputs. View range

RTD temperature meters. View range

Frequently Asked Questions

Pt is the chemical symbol for platinum, and 100 is its resistance in ohms at 0 °C.

It is more accurate, but it costs more and is only specified over a narrower temperature range. Class B is fine for most general process work.

The industrial standard is 0.00385 Ω/Ω/°C. Some older or laboratory sensors use 0.003916, so check the instrument is set to match.

More in This Category

RTD vs Thermocouple vs Thermistor: Which Temperature Sensor Should You Use?

Use an RTD when you need accuracy and long-term stability up to about 600 °C (1,100 °F). Use a thermocouple for high temperatures, fast response or low cost, up to about 1,700 °C (3,100 °F). Use a thermistor for very precise measurement over a narrow, low temperature range, such as in electronics and HVAC.

2-Wire vs 3-Wire vs 4-Wire RTD Connections

The number of wires decides how well the instrument can cancel the resistance of the leads. A 2-wire RTD adds the full lead resistance to the reading, causing errors on anything but very short runs. A 3-wire RTD compensates for lead resistance and is the industrial standard. A 4-wire RTD removes lead effects completely and is used for laboratory and calibration-grade accuracy.

Thermocouple Types (J, K, T, E, N) and Cold Junction Compensation Explained

Each thermocouple type uses a different pair of metals, which sets its temperature range and best uses. Type K is the general-purpose choice up to about 1,260 °C. Type J suits lower temperatures and reducing atmospheres, Type T is best for low and cryogenic temperatures, and Types R and S cover very high temperatures. Every thermocouple measures relative to its cold junction, so the instrument must apply cold junction compensation.

Need help choosing the right product?

Our engineers can help you select, configure and integrate the right sensor, instrument or industrial computer for your application.